Remove the sensor and rinse it thoroughly with tap water. Immerse the sensor in a 0. A fume hood is advised if the HCl is heated. Immerse the sensor in Hamilton storage solution, for a minimum of 60 minutes. Overnight immersion is recommended. Do pH sensors have an expiration date?
Ageing sensors will exhibit the following characteristics: Increased response time to changes in pH Increased impedance of the glass membrane A declining slope value A shift in the asymmetry potential Many of these changes can be corrected for by following the refresh procedure listed above then performing a proper two-point buffer calibration. What speeds up the ageing of the sensor?
What slows the ageing of sensors? My pH sensor is quite old, can it still be used? Related Content Watch a short video with tips on how to increase the lifespan of your pH sensor. The calibration interval needed for pH is highly application-dependent.
The more uniform process conditions are temperature, pressure, composition, freedom from coating, etc. Most pH installations calibrate between once a week and once a month but shorter or longer intervals may be appropriate based on experience.
It is a good practice to begin by calibrating frequently and then gradually go to longer intervals, as stability in the installation and accuracy requirements allow. A change in ORP values for a given process may be due to a number of factors. Instrument electrical stability is seldom a problem. ORP can be influenced by changing mineral composition or changing pH from alternate water sources.
ORP reference electrodes will eventually drift a little with age and will ultimately have to be replaced. Stable ORP requires that a definite oxidizing or reducing condition be present. In pure water, ORP will drift around, responding to levels of dissolved oxygen, traces of contaminants or the last solution in which the sensor was immersed. Thornton does not recommend "calibration" of ORP sensors in solutions.
ORP measured in absolute millivolts from an electrically calibrated instrument will provide a firm basis for comparing readings. Standards for ORP are generally used to verify response of a sensor but their tolerance under industrial conditions is too wide to be used for calibration. Standards are described in ASTM Practice D, including the most commonly used ones which are pH buffer solutions saturated with quinhydrone. Quinhydrone is an organic chemical available from laboratory supply houses.
These solutions must be made up fresh for each use because of air oxidation in storage. The pH and ORP measuring electrode center part of the sensors are glass and platinum respectively and can have very long life as long as they are not broken or chemically attacked.
General purpose pH and ORP sensors also include a reference electrode which has a finite amount of electrolyte. The electrolyte salts must maintain consistent electrical contact with the process sample though the reference junction surrounding the measuring electrode to allow measurement. Because of this contact, it is subject to gradual dissolving and loss in the sample, although this is retarded by a thick gelling agent.
Reference electrode life is highly application dependent, from weeks to years depending on temperature and pressure cycling which expand and contract the electrolyte, tending to "pump" it out of the sensor. In dirty samples the reference junction is also subject to coating and contamination. User manuals can be found in our Resource Library.
Response time should be no longer than 60 seconds and good electrodes should be stable within that time. This response time can depend on the age and cleanliness of your probe and a slow response suggests the electrode is old or dirty. Issues during calibration may be the result of an aged electrode.
Electrodes will generally last months, even if they are not being used. Electrodes will age faster if used in extreme operating conditions strong acids, high temperatures , resulting in a slower electrode response, change in slope, and change in asymmetry i. Rather than displaying pH values, it is best to display mV values on the instrument when checking the sensor performance or calibrating.
Not all instruments have this capability, but some instruments do. Buffer 4 should be to mV away from the buffer 7 mV value, in the positive direction. Buffer 10 should be to mV away from the buffer 7 mV value in the negative direction. If you accept an out-of-range calibration, your probe will not calibrate properly. It is highly likely you will not collect any usable pH data if an out-of-range calibration is accepted.
For refillable electrodes, ensure the fill level of the electrolyte is at least 2 cm above the level of the measurement solution. Replace the electrolyte if it has become contaminated. Depending on the model, the refill opening is either a slider left image or a stopper right image. The refilling opening must always be open during calibration and measurement.
The reference junction must be completely submerged in the solution. The temperature sensor must also be in solution in order to accurately compensate pH for temperature. The pH values of buffer solutions are temperature-dependent as mentioned earlier and the response can vary from manufacturer to manufacturer.
Also, the pH values of buffers in a buffer set can vary from one set to another. Modern pH meters automatically adjust for the respective temperature profile once the buffer set used has been correctly set.
Why is the pH Scale Logarithmic? Anatomy of pH Electrodes. My pH Probe gives static values i. Sorry to hear you're having some issues.
I will pass your email along to our Tech Support department info ysi. I try to calibrate my ph Extech , buffer seven first ok and then buffer four not for good number it is same number 7 how can I fix? Very useful information. So if I were to use a pH meter probe on tissue samples, I should probably clean it with some sort of bleach before using it again? These tips are very useful. I just had a problem with my pH meter, and your tips helped me solve it.
I am really grateful. Are there any solutions that tend to throw a meter out of calibration? High pH or low pH. I had a ex-tech tell me a low pH buffer 4 will throw a meter out of calibration while 7 buffer won't.
Sounds odd to me. This blog series is incredibly useful - not just for me - but for my students - especially the information on pH electrodes - which we all tend to take for granted! By clicking on Submit you agree that Xylem may use your personal data to aid in providing you support, and may contact you directly on this matter. Troubleshooting pH Meter Calibration Problems with 16 Practical Steps Updated June pH is an incredibly important water quality parameter that is measured in nearly every water quality application.
Check the Age of the Probe Probes, or electrodes, for your pH meter generally last months. Perform Routine Maintenance Keeping your pH probe clean can also help eliminate pH calibration problems. Field Electrodes: Regularly soak the electrode in a bleach water solution for minutes to reduce the chances of a clogged reference junction.
Lab Electrodes: Place the electrode for 5 minutes in 0. Check for Physical Damage to the Probe If your probe is damaged, broken glass bulb, crack in the glass, etc. A good way to use reused buffers is for probe rinsing only.